Organelle detection kits

Cells do not appear to be healthy

  • Some cells require serum to remain healthy >> Add serum to the staining solution.  Serum does not affect staining.  Normal amounts of serum added range from 2% to 10%.

Organelles are not sufficiently stained

  • Very low concentration of dye was used or dye was incubated with the cells for an insufficient length of time >> Increase the labeling concentration or increase the time allowed for the dye to accumulate in the organelle.

Blue nuclear counter-stain is too bright compared to the organelle stain

  • Different microscopes, cameras and filters may make some signals appear very bright >> Reduce the concentration of the nuclear counter-stain or shorten the exposure time.

The dyes fail to stain acidic organelles in fixed and/or permeabilized cells.

  • The following kits are only suitable for live cell staining only >> use only for live cell analysis:
    • Cyto-ID Autophagy detection kit
    • Lyso-ID Red or Green lysosomal detection kit
    • Lyso-ID Red Cytotoxicity kit
    • Mito-ID Membrane Potential kits
    • Nuclear-ID Green/Red nucleoli / nuclei detection kit
    • Nucleolar-ID Green nucleoli detection kit
    • Organelle-ID RGB III assay kit
 